Class I Recall for Medtronic SynchroMed and IsoMed Infusion Systems

AddThis Social Bookmark Button

Medtronic Inc.’s January communication regarding problems with its SynchroMed and IsoMed infusion systems has now been deemed a Class I recall by the Food & Drug Administration (FDA).  A Class I recall indicates a situation in which there is a reasonable probability that use of the product will cause injury or death. Medtronic said the classification from the FDA does not change the recommendations made to physicians in the January 2008 letter and there is no new action required of physicians or patients in regards to the defective SynchroMed and IsoMed infusion systems.

Medtronic’s SynchroMed and IsoMed intrathecal drug delivery systems consist of an implantable pump and catheter that deliver small quantities of drug directly into the intrathecal space in the spine.  These devices are used to treat chronic, intractable pain and for the management of severe spasticity of cerebral or spinal origin.

On January 16, Medtronic sent a letter to healthcare providers informing them of an increase in the rate of reported inflammatory mass cases in patients who have received intrathecal drug delivery through its SynchroMed and IsoMed implantable infusion systems.  This letter was an update to two previous communications on this topic that Medtronic issued in 2001 and 2003.  The risk of inflammatory mass formation has been included in the labeling for Medtronic’s SynchroMed and IsoMed implantable drug infusion systems as either a Warning or Precaution, since 2001 and in the prescribing information for Infumorph (preservative-free morphine sulfate for microinfusion pumps) since 2003.

Inflammatory mass is a chronic inflammatory or granulomatous mass at or near the tip of intrathecal catheters and has been reported with the infusion of morphine, baclofen and other physician-prescribed drugs and/or mixtures, including pharmacy-compounded drugs. In the January letter, Medtronic noted an increase in reported cases of inflammatory mass associated with intrathecal drug delivery from 0.1 percent reported to date in 2001 to 0.5 percent reported to date in 2007.  The company said that the actual incidence was likely  higher than stated due to under-reporting.  Medtronic said it has received FDA approval to update the SynchroMed and IsoMed labeling to include this new data.

According to Medtronic’s press release, the most frequently reported symptoms of inflammatory mass are decreased therapeutic response, pain, and neurological deficit/dysfunction.  Serious reported symptoms include paralysis and other neurological impairments.  To date, there have been no reported deaths associated with this issue. Medtronic said the development of inflammatory mass was associated with a wide range of doses and concentrations of opioids.   The risk of inflammatory mass occurrence also appeared to increase over time and with higher concentrations of opioids

In the January letter, Medtronic cautioned clinicians to utilize the lowest effective dose and concentration of opioids and to monitor patients closely for early clinical symptoms of inflammatory mass.  The letter also recommended diagnostic steps for physicians to consider for patients who have new neurological symptoms aimed at preventing more severe outcomes.

Medtronic said patients with questions should talk to their physician or contact Medtronic Patient Services at 1-800-510-6735, Monday – Friday, 8 a.m. to 5 p.m. CDT.  Physicians with medical questions related to Medtronic therapies should contact Medtronic at 1-800-328-0810, Monday – Friday, 8 a.m. to 5 p.m. CDT.

You can leave a response, or trackback from your own site.

« Drug Trials Involving Kids Need More Monitoring Chinese Counterfeiters Suspected in Heparin Contamination »

Leave a Reply




NUMBNESS ARMS LEGS- TINGLING DENTURES- MUSCLE WEAKNESS?

Do you have these symptoms and use denture cream? You may have serious side effects? Get the facts! DENTURECREAMLAWYER.COM

DENTURE CREAM NEUROPATHY

Super Poligrip or Fixodent May Cause Nerve Damage. 800-LAW-INFO NEUROPATHY-DENTAL-CREAM.COM

Victim of SECURITIES FRAUD?

Many people have lost money due to securities fraud.
Are you one of them? Get the facts! SECURITIES-FRAUD.COM

Injured By CIPRO?

CIPRO Has Been Found To Cause Tendon Damage and Tendon Ruptures.
Were You Harmed By This Drug? CIPRO-TALK.COM

Gardasil - Learn The Facts

Gardasil Has Been Called The Wonder Drug To Prevent Cervical Cancer. See The Facts
www.gardasil-talk.com

Paxil Can Cause Birth Defects

If your child was born with a Birth Defect While You Were On Paxil, We Can Help
YourLawyer.Com

Paxil Birth Defect and Side Effects Info

Get all of the info you need regarding Paxil and it serious side effects PAXIL-TALK.COM

Nuva Ring

The NuvaRing can cause Blood Clots, Strokes and more. www.yourlawyer.com.com

Whistle Blower

Stand up and say what you think is right. See something, say something. www.whistlebloweradvisor.com

Gadolinium MRI Contrast

NSF & NFD From MRI or MRA? Call Our Attorneys Today www.gadolinium-mri.com